GASS-2040 Sampling System
The GASS-2040 conditions high-flow, high-moisture
samples, eliminating acid mists or ammonia if present. It processes the
toughest samples, with flows of up to 25 liters per minute and moisture
contents in excess of 50%. The GASS-2040 can be mounted anywhere,
including on the stack flange with an integral stack probe, which
eliminates heated sample lines.
The GASS-2040 System Design
The GASS-2040 contains three
temperature-controlled zones mounted in an environmentally sealed,
NEMA-4X housing. In the first zone, the sample passes through a heat
exchanger, then through a coalescing particulate filter to remove
particles down to 0.1 micron. Acid mists, if present, are coalesced then
removed by an automatic drain. If ammonia is present, a proprietary
ammonia scrubber removes it.
In the second zone, the sample passes
through a proprietary Nafion® dryer. The initial portion of the
dryer is heated above the sample dew point to prevent condensation and
to create greater water vapor removal.
In the third zone, the sample passes
through the remainder of the dryer, further reducing the dew point down
as low as -25°C, depending upon dryer model selected and sample flow
rate. The system PLC with optional Ethernet connection for remote and
flow controls for the dryer purge gas. Optional accessories, if
required, such as a sample pump or pressure-swing air dryer for the
purge air are also mounted here. This zone operates at ambient
temperature so these controls and accessories do not overheat.
When an integral sampling probe is added,
the GASS-2040 can be mounted right on the stack flange. The probe filter
and blowback assembly are mounted inside the GASS-2040, and the system
PLC controls filter temperature and blowback timing. Conditioning the
sample immediately after it leaves the stack improves its analytical
calibration times, saving calibration gases.
|