AS-Series Media Replacement

Replacing the Scrubbing Media:

Ammonia Scrubber Media Life Curve

  1. If you have a heated version, unplug the device and remove the heater blanket

  2. Unscrew the thumbscrew on the bottom of the housing

  3. Rotate the yoke which holds the housing together

  4. Remove the old media and dispose of properly.

  5. Rinse the housing with soapy water to clean

  6. Fill the housing with 150cc of part A material (Ceramic burl saddles)

  7. Fill the housing with 50cc of part B material (White pellets)

  8. Replace housing on top manifold

  9. Turn housing gently to allow o-ring on center tube and shell tube to seat properly

  10. Replace the yoke and tighten thumbscrew finger tight. Do not over-tighten! Over tightening will cause splintering of the plastic shell.