Start by removing the existing
yoke assembly.
-
Unscrew the thumb screw at the
bottom of the ammonia scrubber (AS) housing.
-
Place the thumb screw to the
side for use on the new yoke assembly.
-
Using a 1/8" allen wrench,
remove both shoulder screws threaded through the yoke arm.
-
Remove the existing yoke arm.
-
Place two of the plastic
washers, used on the existing yoke assembly to the side for
use on the new yoke.
-
Attach the supplied heater
jacket around the AS shell using the hood and loop fastener
strip.
-
Install the new yoke arm by
placing one of the new spacers on the inside of the yoke arm
between the AS top and the yoke arm.
-
Place one of the plastic
washers, set aside from step 6, on the outside of the new
yoke arm.
-
Thread a new shoulder bolt
through the plastic washer, yoke arm hole and spacer into
the AS top.
-
Repeat steps 8 thru 10 for the
other side.
-
Reattach the thumb screw
through the bottom of the new yoke arm.
-
Connect 120/240 VAC power
supply lines to the power supply lines on the heater jacket.
-
Allow heater jacket to run for
approx. 15-30 minutes before starting sample flow.
-
The heater is equipped with a
thermocouple and thermostat to regulate the power
requirements of the heater.