Baldwin™- Series SO3 Aerosol Removal System

Specifically Designed to Remove SO3 Aerosol
The Perma Pure Baldwin-Series Aerosol removal system is specifically
designed for gas sample applications with relatively high SO3 content
(>10ppm). The SO3 aerosol removal series has special timing circuits
that alternate freeze and thaw the precisely packed heat exchangers.
Before an active heat exchanger thaws, the alternate active heat
exchanger reaches the -7C set point to ensure uninterrupted operation
and no SO3 slippage. This sub zero temperature forms a thin ice layer
that captures and removes SO3 aerosol from the sample stream.
Operation
The two heat exchangers located on the left side of the sample cooler
are connected in series. The sample stream first passes through an
inactive exchanger used for removing gross amounts of water. The sample
then passes to an active glass heat exchanger. The exit dew
point from this heat exchanger is controlled to +4°C, thereby reducing he
moisture concentration to less than 1%. The gas sample then flows
through the gas sample pump where it alternately passes through frozen Kynar® packed heat exchangers, located on the right side of the cooler,
then to three-way Teflon® solenoid control valve and finally to the
remainder gas sampling system. Condensate is pumped from each heat
exchanger by a dedicated peristaltic drain pump head.
|
Model 10410 |
|
Common Industry Applications |
Coal Combustion
Glass Furnaces
Refineries |
|
Standard Sample Gas Flow Rate |
1-5 LPM
|
|
# of Impingers |
2 Glass 10"
2 Packed Kynar 10" |
|
Dimensions |
14.5x12.75x12.25 in. HWD |
| Weight |
36 lbs |
Available Options: Acid Safety Scrubber Kit (Part No.
3SCR-004); Water Slip Sensor w/Holder Assembly (Part no. 3KCW-002) Note:
Water slip sensor requires 3CCB-018 Alarm board. |