MG-1228 Probe Specifications

Gas Analysis Sampling System with Integral Probe

Mini-GASS™ Probe systems eliminate heated lines entirely by mounting the conditioning system directly on a stack probe. The sample is dried and cooled immediately after removal from the stack to minimize changes in sample composition. This system eliminates the expense of installation, operation, and maintenance of heated lines, while faster sample equilibration reduces calibration times, saving on the expense of calibration gases.

A series of filters conditions the gas stream before the sample is dried in the Perma Pure Nafion membrane dryer. The Mini-GASS Probe removes particles, acid mists, and water vapor without analyte loss due to the selectivity of the dryer.

The Mini-GASS Probe incorporates all the features of the Mini-GASS system and can be configured with the same options. The Mini-GASS Probe also offers an optional heated blowback assembly for its filter to minimize maintenance with high particle levels, and Hastelloy® probe components for additional corrosion resistance and temperature tolerance.

Dimensions 12.63" x 34.75" x 7"
Weight 35 Lbs.
System Power Connection 120VAC, 5A buss type
Alarm Relays 6 pole terminal block provided for the connection of external alarm equipment. Contacts rated for 250VAC, 3A resistive.
  1. Low Purge - Flow < 10 lpm
  2. Low Temperature - Temp > 4 degrees below the set point temperature
Calibration gas requirement Sample flow rate plus 10%, 1/4" tube fitting
Enclosure NEMA 4X, Fiberglass
Heatless Dryer Optional, Maximum pressure 100 psig, oil-free non-condensing compressed air
Blowback Tank Volume 0.7 scf when at 100 psig
Duration One half second minimum
Heater, Optional 125 °C
Connection 3/8" tube fitting
Probe Filter Power requirement 150 Watts at 115 or 230 VAC
Set temperature 350 F (180 C)
Element 2 micron ceramic

 

MG-1228 stack mounted sampling system

Mini-GASS Probe