Micro-GASS Temperature Controller
Cal 3300 Temperature Controller Instructions
-
For all Micro-GASS
systems with the CAL3300 Temperature controller, perform the
following setup procedures.
-
Power Up
- Apply power to the system and wait for the controller to
perform its self test routine. The display will then be
alternately flashing "nonE" and "inPt".
Input sensor type
- Press and hold * key. Press up key repeatedly to select the "tc
K" option for K type thermocouple. Release both keys.
Input unit type
- Press the up key once. The display will then be alternately
flashing "unit" and "nonE". Press and hold the *
key. Press up key repeatedly to select the "OC"
unit. Release both keys.
Input set point 1 type
- Press the up key once. The display will then be alternately
flashing "SP1.d" and "nonE". Press and hold the *
key. Press up key once to select the output device "SSd".
Release both keys.
Input the maximum
temperature set point - Press and
hold the down key until "LEVL" appears. Press and hold
the * key. Press the down key to move to level 2. Release the
two keys. Press and hold the up key until "hi.SC"
appears. Press and hold the * key. Press the down key until "65"
appears. Release the two keys.
Input the minimum
temperature set point Press the up
key once. "Lo.SC" appears. Press and hold the * key.
Press the up key until "45" appears. Release the two
keys.
-
Configure the low
temperature alarm.
-
Configure SP2 relay as alarm
- Press and hold up and down keys
simultaneously for 3 seconds to enter the program mode. Release
the two keys. The display will then be alternately flashing "tunE"
and "oFF". Press the down key once. The display will then
be alternately flashing "LEVL" and "1". Press and
hold the * key. Press the up key once to
move to level 2. Release the two keys. Press and hold the
up key until "SP2.A" appears. Press and hold the * key.
Press the up key until the display reads FS.Lo. Release
the two keys. The display will then be alternately flashing "SP2.A"
and "FS.Lo".
Save configuration into
memory - Press and hold up and
down keys simultaneously for 3 seconds. The display will then be
alternately flashing the process temperature and "PArK"
indicating that no set point temperature has been entered.
Set the control set point and
auto-tune the controller .
-
Input the control set point
temperature - Press and hold the *
key. Hold the up key until the display goes from 0 to
55.
-
Auto-tune the controller -
Press and hold up and down keys
simultaneously for 3 seconds to enter the program mode. Release
the two keys. The display will then be alternately flashing "tunE"
and "oFF. Press and hold the * key. Press the up key
until "At.SP" appears. Release the two keys. Press and
hold up and down keys simultaneously for 3 seconds to exit the
program mode. The display will then be alternately flashing "tunE",
"At.SP" and the actual process temperature. Allow the
controller time to auto-tune. When tuning is complete the
display will show only the actual process temperature.
This process may take up to an hour.
Enable low temperature alarm.
Input the alarm set point
- Press and hold the down key until "LEVL" appears. Press
and hold the * key. Press the down key to move to level 1.
Release the two keys. Press and hold the up key until "Set.2"
appears. Press and hold the * key. Press the up key until "40"
appears. Release the two keys. This alarm will control the
sample pump so that it will automatically turn off at any time
that the temperature is below 40OC.
|