Micro-GASS Setup Instructions

Installation


Enclosure Mounting

The Micro-GASS enclosure is rated NEMA 4, but the system is intended for indoor installation. For mounting, choose a location to avoid exposure to direct sunlight as it may raise the internal temperature above the operating limits, nor should the system be subjected to extreme cold (<10C). Vertical orientation of the system is required for proper operation of the flow meter. A set of four mounting feet are supplied for mounting the system on 12-1/4" centers.

 

Sample Connections

Sample Inlet and Heated Line Connection - To prevent condensation of the sample gas a heated line may be required. The Micro-GASS is supplied with a heat shrink sealing heated-line fitting sleeve or "boot". Connect the boot to the enclosure with the fitting hub inside the enclosure and the o-ring seal on the outside, as shown in the figure, tighten to secure. The heated sample line passes through the sleeve into the enclosure. Remove the insulation from the heated line and connect the sample line to the 1/4" compression fitting at the inlet to the filter body. Heat shrink the boot sleeve to the heated line:Micro-GASS Inlet fitting connector

  1. Before heat shrinking the boot connect the sample line to the filter and ensure that it is in its final and ultimate position. The sleeve is adhesive lined and the heated line cannot be removed after shrinking.

  2. Provide adequate ventilation. Fumes from burning can cause eye, skin, nose and throat irritation. Using a low intensity heat source; such as a hot air heat gun, or infrared heat source, heat the boot evenly until the sleeve shrinks into place.

  3. Allow extra heating time to ensure that the adhesive melts and flows for proper adhesion. Keep the heat source 1" away from the heated line.

  4. Keep the heat source moving; do not concentrated the heat in one place. If charring or burning does occur, stop heating immediately.

Dry Sample and Purge Connections - 1/4" Bulkhead compression fittings located on the side of the enclosure are for theses connections. Connect the "Dry Sample Outlet" port on the Micro-GASS enclosure to the inlet port of the sensor. Connect hte sensor exhaust/outlet back to the "Dry Sample Return" on the Micro-GASS. No heated line is required for these connections unless the lines will be exposed to cool temperatures (<10°F).

Sample Exhaust - The sample exhaust can either vent freely, or can be piped away from the Micro-GASS. This gas is "wet" and condensation in the line is likely.

Calibration Gas Connection – ¼” Bulkhead compression fitting is located on the enclosure bottom. The port must be plugged when not in use, a cap is provided.

Electrical Connections - 60 Watts, 110 or 220 Volts

Standard 110VAC receptacle is provided. For 220-volt systems, a pigtail cord is provided for connection to the appropriate 220-volt plug (not provided).

 

Heated Line – For connection to the heated line two female and one male bullet connectors; for neutral, line, and ground connections respectively, are provided. An in-line fuse holder is provided for a fuse not to exceed 10 amps.