PD-Series Home
Performance
Setup
PD Cleaning
Brochure
Manual
Molded Dryer Manual

PD-Series Gas Dryers Setup

The Perma Pure PD™-Series gas dryers are shell and tube moisture exchangers that allow the transfer of water vapor between two countercurrent flowing gas streams. The devices consist of an inner tube of Nafion® polymer surrounded by an outer tube of various inert materials. The water vapor is selectively absorbed into the walls of the inner Nafion tubes and transferred to a purge gas stream. This transfer is driven by the difference in the partial pressures of the water vapor on opposing sides.

PD-Series gas dryer schematic

Typically the sample flows through the inner tubes and the purge gas flows in the opposite direction within the shell. The purge gas stream should be -40°C dew point air, nitrogen or other gas.

The dry gas entering the dryer at the sample outlet performs two functions:

  1. Provides a medium to carry away water vapor from the sample.

  2. Creates a temperature gradient along the length of the dryer. For more information see Temperature Effects on Nafion.

When installing the PD-Series gas dryer, the following rules apply:

  1. The sample pressure must be higher than the purge pressure to prevent the tubing from collapse. Note that the dryer may be operated with the sample under vacuum as long as a larger vacuum is pulled on the purge (shell) side. 

  2. The inlet third of the dryer must be maintained at 5 to 10° above the sample dew point to prevent condensation. 

There are several ways to operate the PD-Series gas dryers.

  1. Supply dry purge air from a desiccant air dryer or nitrogen. The flow should be 2-3 times the sample flow rate. This is the most efficient method. If no dry purge gas is available, one of following methods may be used.

  2. The reflux setup returns the dry sample back to the dryer for use as the purge after it has gone through the analyzer. Since the reflux method uses all of the sample as the purge gas, only the sample flow required for the analysis passes through the dryer. This results in a very high drying efficiency. The vacuum on the purge gas should be at least 15" of Hg, with a higher vacuum preferable. This vacuum level is necessary to provide the desired 2:1 purge-to-sample flow ration based on the actual volumetric flow.

PD-Series Reflux Setup Schematic

  1. The split sample method diverts some of the sample from the main stream to be used as the purge gas. As a result, more sample passes through the dryer than is required for the analysis and somewhat lower drying efficiency will occur. The flow diagrams below shows how to operate with the sample under positive pressure, setup A or under vacuum, setup B.

To determine the purge flow rate required for the split sample method, the following equation can be used.

Vp = Vs / [(Ps/2Pv)-1]

Where: Vp = Purge flow rate, Vs = Sample flow rate, Ps = Sample pressure (in absolute units), Pv = Purge pressure (in absolute units). Any units may be used as long as they are consistent.

PD-Series Split Stream Setup

For a print version, click on pdf manual.

PD-Series Home Performance Setup PD Cleaning Brochure Manual Molded Dryer Manual