Nafion Dryer Purge Gas Method #4 – Split Sample Gas with Vacuum

6 views 1

  • Most popular method used by Perma Pure customers by unit volume
  • The split sample method is similar to the Reflux Method previously described except that it channels a portion of the sample gas through the Nafion tubing serving as the purge gas before the analyzer.
  • Choose the split sample method if it is not possible or practical to route the analyzer exhaust back to the gas dryer.
  • The dryer must be up-sized for the increased flow rate adding the portion used for the purge gas to the sample gas flow rate required for the analyzer.
  • Additionally, the right purge gas amount will need to be selected to provide optimum drying performance.
  • Provides a consistent dryness level at the output when run continuously.